P2.8 Looking for filter defects/artifacts + Science detector calibrations repeat
- Phase : 2
- Status : Draft
- Category : Daytime testing (functionality and operability) (C2)
- Recurrence : once at commissioning
- Duration : 2 days
Description
- Finding the filter defects/artifacts using an internal calibration source (flat /on-axis PSF?) within the instrument.
Prerequisites
- NCPA measurement (within SHARK-NIR bench using internal sources) available.
Procedure
- Take the images of all the filters using internal calibration source lamps within the instrument to locate/find out the filter effects/artifacts.
- You may take short- and long-exposures to get better contrast of the features if required.
- Repeat the science calibrations as done in P1.9.
- Perform the working of this monitoring activity Filter wheels reproducibility monitoring.
Success criteria
- All the filter images saved.
Notes
- The time dedicated to this activity has to be considered in combination with the next activity if needed. Parallelism is possible.
LBTO support
Associated SHARK-NIR personnel
Date performed and by whom
Comments
MB: I wonder why this needs to happen in this phase. I would say it could be done at any moment with internal sources, except of course for the dichroic / LBT mirrors