Version 1 (modified by kradhakrishnan, 4 years ago) (diff) |
---|
SDC6 Detector persistence
- Phase : Any
- Status : Draft
- Category : Daytime calibration (C3)
- Recurrence : once every 1-3 months starting from commissioning.
- Duration : 20 minutes
Description
- Monitor persistence after saturation.
Prerequisites
- Instrument and dome closed. No lights inside dome.
Procedure
- Record a background exposure, then one highly saturated exposure; then a long series of exposures with source off.
- CI B_H, point source.
Success criteria
- Accuracy of photometry 10%
Notes
- Same as activity 9.3 in AD4
LBTO support
Associated SHARK-NIR personnel
Date performed and by whom