Changes between Version 2 and Version 3 of docs/CommissioningPlan/SDC6


Ignore:
Timestamp:
01/14/22 13:15:22 (3 years ago)
Author:
kradhakrishnan
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • docs/CommissioningPlan/SDC6

    v2 v3  
    11== SDC6 Detector persistence ==
    2 •       **Phase**               :       Any 
    3 •       **Status**              :       Draft   
    4 •       **Category**            :       Daytime calibration (C3)
    5 •       **Recurrence**  :       once every 1-3 months starting from commissioning.
    6 •       **Duration**            :       20 minutes
     2
     3- **Phase**             :       Any 
     4- **Status**            :       Draft   
     5- **Category**  :       Daytime calibration (C3)
     6- **Recurrence**        :       once every 1-3 months starting from commissioning.
     7- **Duration**  :       20 minutes
     8
    79**Description**
    8 •       Monitor persistence after saturation.
     10- Monitor persistence after saturation.
     11
    912**Prerequisites**
    10 •       Instrument and dome closed. No lights inside dome.
     13- Instrument and dome closed. No lights inside the dome.
     14
    1115**Procedure**
    12 •       Record a background exposure, then one highly saturated exposure; then a long series of exposures with source off.
    13 •       CI B_H, point source.
     16- Record a background exposure, then one highly saturated exposure; then a long series of exposures with source off.
     17- CI B_H point source.
     18
     19**Templates**
     20
    1421**Success criteria**
    15 •       Accuracy of photometry 10% 
     22- Accuracy of photometry 10%.
     23
    1624**Notes**
    17 •       Same as activity 9.3 in AD4
     25- Same as activity 9.3 in AD4
     26
    1827**LBTO support**
    1928
    20 **Associated SHARK-NIR personnel**
    21 
    22 **Date performed and by whom**,
    23 
    24 
    25 **Comments**
    26 
    27 MB: Also this might happen only in clean room
    28 Concerning teh duration it should be better specificed until how much time after the saturated exposure you want to keep recording images. This will drive the duration of the tests
     29**Date performed and by whom**