Version 5 (modified by kradhakrishnan, 3 years ago) (diff) |
---|
SDC4 Thermal background
- Phase : Any
- Status : Draft
- Category : Daytime calibration (C3)
- Recurrence : once every month starting from commissioning.
- Duration : 1 hour(s) (this duration is valid only during regular operations, i.e., after commissioning)
Description
- Measure the instrument background and seasonal variations.
Prerequisites
Procedure
- This calibration is performed by reading the detector without any light
- CI BB_H, DIT=1800s, NDIT=2.
Templates
- SHARKNIR_gen_mon_ThermalBackground (not released)
Success criteria
- Accuracy of 1.2 e-/pix (FWHM of the distribution of a difference of backgrounds).
Notes
- Same as activity 9.1 in AD4
LBTO support
Date performed and by whom