P3.5 Orient the coronagraphs
- Phase : 3
- Status : Draft
- Category : Nighttime testing (functionality and operability) (C4)
- Recurrence : once at commissioning
- Duration : 1.5 hour
Description
- Adjust the orientation of the coronagraphic masks with respect to the true orientation of the M2 spider by rotating the SHARK-NIR derotator.
Prerequisites
- Atmospheric conditions: should be able to see bright stars in the sky (preferably close to zenith?).
- Authorize SHARK-NIR on the SX side.
- Suitable bright enough star close to zenith to have negligible effects from flexures (including the possibility of performing AO).
Procedure
- Pointing to a star, the first coronagraph is inserted in the path. The bearing is rotated maximizing the contrast achieved on the SCICAM. (use the pupil lens to see the pupils alignment maybe)?
- This operation is repeated for all the coronagraphs.
Success criteria
- All the coronagraph are aligned to the true orientation of the M2 spider by maximizing the contrast at SCICAM.
- Note down the contrast achieved for all the coronagraphs including the atmospheric conditions and the brightness of the on-axis star.
Notes LBTO support
- SOUL-AO support
Associated SHARK-NIR personnel
Date performed and by whom Comments
MB: when saying orientation better to specify through wheel rotation DONE.