19 | | * Check the phase diversity NCPA approach using Wollaston prism + DI-DBI (need to do on the spot, but script exists). Take images at SCICAM. |
20 | | * Check the phase diversity approach using the internal calibration source (already calibrated). Take images at SCICAM. |
| 19 | * Check the phase diversity NCPA approach using Wollaston prism + DI-DBI (already calibrated). Take images at SCICAM. |
| 20 | * Measure the phase diversity NCPA approach using Wollaston prism + DI-DBI and using real star (need to do on the spot, but script exists). Take images at SCICAM. |
| 21 | * Digression: If the PSF after applying the calibrated NCPA is bad, we could close the SHARK-NIR shutter (can have the SOUL-AO running parallel) and do a quick internal NCPA measurement (either till the NCPA modal coefficients or until the new shape on ALPAO-DM). Confirm that the alignment of SHARK-NIR is still good. |