Changes between Version 18 and Version 19 of docs/CommissioningPlan/P2.8


Ignore:
Timestamp:
01/14/22 17:58:24 (3 years ago)
Author:
kradhakrishnan
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • docs/CommissioningPlan/P2.8

    v18 v19  
    1  == P2.8        Redo science detector calibrations ==
    2 •       **Phase**               :       2 
    3 •       **Status**              :       Draft   
    4 •       **Category**            :       Daytime testing (functionality and operability) (C3) + Daytime calibration (C4)
    5 •       **Recurrence**  :       once at commissioning
    6 •       **Duration**            :       1 day
     1 == P2.8 Checking pointing offset between SHARK-NIR and SOUL-AO==
     2 
     3- **Phase**             :       2 
     4- **Status**            :       Draft   
     5- **Category**  :       Daytime testing (functionality and operability) (C3) + Daytime calibration (C4)
     6- **Recurrence**        :       once at commissioning
     7- **Duration**  :       1 day
     8
    79**Description**
    8 •       After the installation of the SHARK-NIR at the LBT platform, the science detector is tested for all its read-out modes, windowing possibilities. Also, some of the calibration scripts are rechecked.
     10- Test to check if there is a considerable pointing offset between the SHARK-NIR (SCICAM/TECCAM) and SOUL-AO as a function of the elevation of the telescope
     11
    912**Prerequisites**
    10 •       SCICAM cooled down.
     13- RR is mounted on the SX side.
    1114
    1215**Procedure**
    13 •       Repeat the science calibrations below (some of these are performed when SHARK-NIR was in the LBT cleanroom or at Padova cleanroom. Check [wiki:/docs/CommissioningPlan/P1.11 P1.11]) for details.
    14 ||**Code**||**Activity**||**Location to perform this test**||
    15 ||SDC1||[wiki:/docs/CommissioningPlan/SDC1 Instrumental background for science]||Padova / LBT cleanroom / LBT Platform||
    16 ||SDC2||[wiki:/docs/CommissioningPlan/SDC2 Instrument flat field, detector linearity and bad pixels]||Padova / LBT cleanroom||
    17 ||SDC3||[wiki:/docs/CommissioningPlan/SDC3 Wavelength calibration]||Padova / LBT cleanroom?||
    18 ||SDC4||[wiki:/docs/CommissioningPlan/SDC4 Thermal background]||Padova / LBT cleanroom / LBT Platform||
    19 ||SDC5||[wiki:/docs/CommissioningPlan/SDC5 Detector gain]||Padova / LBT cleanroom / LBT Platform?||
    20 ||SDC6||[wiki:/docs/CommissioningPlan/SDC6 Detector persistence]||Padova / LBT cleanroom / LBT Platform?||
    21 ||SDC7||[wiki:/docs/CommissioningPlan/SDC7 Spectral resolution]||Padova / LBT cleanroom||
    22 ||SDC8||[wiki:/docs/CommissioningPlan/SDC8 Ghost calibration]||Padova||
     16- TBD
     17
     18**Templates**
    2319
    2420**Success criteria**
    25 •       All calibrations successfully tested.
    26 * All relevant information (files) are saved in the repository.
     21- Estimate the pointing offset and conclude if it is acceptable or not. If not acceptable, take proper actions to overcome this issue.
     22
    2723**Notes**
    28 •       The time dedicated to this activity has to be considered in combination with other activities. Parallelism is possible.
     24
    2925**LBTO support**
    30 
    31 **Associated SHARK-NIR personnel**
     26- Authorize SHARK-NIR on the SX side.
     27- RR is mounted on the SX side.
     28- AO support
    3229
    3330**Date performed and by whom**
    34 