Version 5 (modified by kradhakrishnan, 4 years ago) (diff) |
---|
P2.8 Looking for filter defects/effects/features
- Phase : 2
- Status : Draft
- Category : Daytime testing (functionality and operability) (C2)
- Recurrence : once at commissioning
- Duration : 2 days
Description
- Finding the filter defects/effects/features using different lamps within the instrument.
- If required, this may be repeated for on-axis source from RR for selected or all filters while the AO loop is running (daytime)
Prerequisites
- If relevant, estimate and remove the NCPA using phase-diversity method.
- RR mounted on SX side.
Procedure
- Take the images of all the filters using different lamps within the instrument to locate/find out the filter effects/defects/features.
- You may take short- and long-exposures to get better contrast of the features if required.
- This may be repeated with the on-axis source from the RR with AO running to check for any extra features arriving from outside SHARK-NIR but from LBT mirrors or the deployable arm.
- Repeat the science calibrations as done in P1.9.
- Perform the working of this monitoring activity Filter wheels reproducibility monitoring.
Success criteria
- All the filter images saved.
Notes
- The time dedicated to this activity has to be considered in combination with the next activity, if needed. Parallelism is possible.
LBTO support
- Authorize SHARK-NIR on SX side.
- RR mounted on the SX side.
- Support from LBT SW Team, LBTI/Arcetri team to run SOUL AO
Associated SHARK-NIR personnel
Date performed and by whom