  • docs/CommissioningPlan/P1.8

    v16 v17  
    1616
    1717**Procedure**
    18 - Use the on-axis calibration fiber and check if the PSF position on the scientific detector is consistent with the one recorded before shipping for multiple bearing angles. Perform Phase Diversity measurement to compare the retrieved aberrations on the on-axis PSF with the ones measured before shipping.
     18- Use the on-axis calibration fiber, check if the PSF position on the scientific detector is consistent with the one recorded before shipping for multiple bearing angles. Perform Phase Diversity measurement to compare the retrieved aberrations on the on-axis PSF with the ones measured before shipping.
    1919- Check the pupil position at the detector and verify the consistency with the one before shipping.
    2020- Check the relative alignment of the coronographic masks.
     
    3535
    3636**LBTO support**
    37 - Liquid Nitrogen refill is required in the cleanroom
     37- Liquid Nitrogen refill is required in the cleanroom.
    3838
    3939**Date performed and by whom**