19 | | - When the services are up and running, after the motors are sent to home positions, each of the individual motors is individually tested for their different named/nominal positions. |
20 | | - The light sources within SHARK-NIR are used simultaneously to check the working of the motors looking at the optical feedback. |
21 | | - Check the CRED-2 and ALPAO-DM SW using optical feedback. |
22 | | - Take the images of all the filters using the internal calibration source (flat-field) within the instrument to locate/find out the filter defects/artefacts, if any. You may take short- and long-exposures to get better contrast of the features if required. |
23 | | - Check the filter wheel position monitoring script detailed in [wiki:/docs/Commissioning/SNM1 SNM1]. Save the data in the repository. |
| 20 | - Check bias calibration & master dark scripts and then perform the following daytime calibration procedures below. |
| 21 | |
| 22 | ||**Code**||**Activity**||**Location to perform this test**|| |
| 23 | ||SDC1||[wiki:/docs/CommissioningPlan/SDC1 Instrumental background for science]||Padova / LBT cleanroom / LBT Platform|| |
| 24 | ||SDC2||[wiki:/docs/CommissioningPlan/SDC2 Instrument flat field, detector linearity and bad pixels]||Padova / LBT cleanroom|| |
| 25 | ||SDC3||[wiki:/docs/CommissioningPlan/SDC3 Wavelength calibration]||Padova / LBT cleanroom|| |
| 26 | ||SDC4||[wiki:/docs/CommissioningPlan/SDC4 Thermal background]||Padova / LBT cleanroom / LBT Platform|| |
| 27 | ||SDC5||[wiki:/docs/CommissioningPlan/SDC5 Detector gain]||Padova / LBT cleanroom / LBT Platform|| |
| 28 | ||SDC6||[wiki:/docs/CommissioningPlan/SDC6 Detector persistence]||Padova / LBT cleanroom / LBT Platform|| |
| 29 | ||SDC7||[wiki:/docs/CommissioningPlan/SDC7 Spectral resolution]||Padova / LBT cleanroom|| |
| 30 | ||SDC8||[wiki:/docs/CommissioningPlan/SDC8 Ghost calibration]||Padova|| |
| 31 | |
| 32 | - Check the working of all the detector read-out modes and windowing possibilities. Note that this will be checked rigorously in the Padova laboratory. |
26 | | - SHARKNIR_ci_mon_LyotMaskPos (not released) |
27 | | - SHARKNIR_di_mon_MaskRepeatability (not released) |
| 35 | - SHARKNIR_cal_InstrumentBackground (not released) |
| 36 | - SHARKNIR_gen_cal_InstrumentFlatField (not released) |
| 37 | - SHARKNIR_lss_cal_Wavelenght (not released) |
| 38 | - SHARKNIR_gen_mon_ThermalBackground (not released) |
| 39 | - SHARKNIR_di_mon_DetectorGain (not released) |
| 40 | - SHARKNIR_ci_mon_DetectorPersistence (not released) |
| 41 | - SHARKNIR_lss_mon_SpectraResolution (not released) |
| 42 | - SHARKNIR_cal_mon_Ghost (not released) |
30 | | - All the motors can reach the named positions / nominal positions. |
31 | | - All the light sources are working and confirmed through optical feedback at the SCICAM. |
32 | | - Filter wheel position monitoring script is run successfully and data is saved in the repository. |
33 | | - CRED-2 AND ALPAO-DM working normally. |
34 | | - All the filter images are saved. |
| 45 | - SCICAM linearity as expected. |
| 46 | - SCICAM bad pixels mask created. |
| 47 | - SCICAM bias and flat scripts working. |
| 48 | - All individual activity (in the table) success criteria are met. |