Changes between Version 13 and Version 14 of docs/CommissioningPlan/P1.5


Ignore:
Timestamp:
01/28/21 15:45:56 (4 years ago)
Author:
kradhakrishnan
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • docs/CommissioningPlan/P1.5

    v13 v14  
    1919* Check if the SW (HW if present and need to be checked) limit switches are triggered and check the logs.
    2020* Check the electronics/SW of the CRED-2 and ALPAO-DM. Make sure that they both can be individually switched on and off remotely.
     21* Start the SCICAM electronics and acquire (and store) an image in warm conditions.
    2122
    2223**Success criteria**
     
    3435
    3536**Comments**
    36 LM: add in the procedure also starting of SCICAM electronics and try to acquire and store an image (in warm conditions of course)
     37LM: add in the procedure also starting of SCICAM electronics and try to acquire and store an image (in warm conditions of course) [DONE]