P1.11 Check the working of the internal NCPA routine and looking for filter defects/artifacts

  • Phase : 2
  • Status : Draft
  • Category : Unpack, Assemble, and Verification (C1)???
  • Recurrence : once at commissioning
  • Duration : 1 day

Description

  • Check the working of the internal NCPA routine. Finding the filter defects/artifacts using an internal calibration source (flat-field) within the instrument.

Prerequisites

  • SCICAM cooled down.

Procedure

  • Run the Phase-diversity routine using the images taken with the SHARK-NIR internal calibration sources to measure the NCPA within (excluding the deployable arm)
  • Take the images of all the filters using the internal calibration source (flat-field) within the instrument to locate/find out the filter defects/artifacts, if any.
  • You may take short- and long-exposures to get better contrast of the features if required.

Success criteria

  • Internal NCPA routine works fine and the PSF shows improvement in its quality.
  • All the filter images saved.

Notes

  • The time dedicated to this activity has to be considered in combination with the other activities if needed. Parallelism is possible.

LBTO support

Associated SHARK-NIR personnel

Date performed and by whom

Comments

  • KR: add the filter monitoring routine check.