P1.11 Check the working of the internal NCPA routine and looking for filter defects/artifacts
- Phase : 2
- Status : Draft
- Category : Unpack, Assemble, and Verification (C1)???
- Recurrence : once at commissioning
- Duration : 1 day
Description
- Check the working of the internal NCPA routine. Finding the filter defects/artifacts using an internal calibration source (flat-field) within the instrument.
Prerequisites
- SCICAM cooled down.
Procedure
- Run the Phase-diversity routine using the images taken with the SHARK-NIR internal calibration sources to measure the NCPA within (excluding the deployable arm)
- Take the images of all the filters using the internal calibration source (flat-field) within the instrument to locate/find out the filter defects/artifacts, if any.
- You may take short- and long-exposures to get better contrast of the features if required.
Success criteria
- Internal NCPA routine works fine and the PSF shows improvement in its quality.
- All the filter images saved.
Notes
- The time dedicated to this activity has to be considered in combination with the other activities if needed. Parallelism is possible.
LBTO support
Associated SHARK-NIR personnel
Date performed and by whom
Comments
- KR: add the filter monitoring routine check.