Changes between Version 1 and Version 2 of docs/CommissioningPlan/SDC8


Ignore:
Timestamp:
01/29/21 15:53:43 (4 years ago)
Author:
mbergomi
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • docs/CommissioningPlan/SDC8

    v1 v2  
    2020
    2121**Date performed and by whom**
     22
     23
     24**Comments**
     25MB: this is happening with a closed instrument, correct? If so, it should be done in Padova the first time, then, if needed, checked and or re-taken either in clean room or while on the platform.
     26
     27MB: The timing seems high. Essentially this activity only imples the recording of images at different configurations. Putting 5 minutes per image should be enough. Also, it would be useful to put the number of configurations.
     28
     29MB: comments for scientists. From the technical point of view you are requesting us to acquire these images. Are you then analyzing them and reporting somewhere the location and intensity of found ghosts fro different configurations?
     30
     31
     32