Changes between Initial Version and Version 1 of docs/CommissioningPlan/SDC4


Ignore:
Timestamp:
01/15/21 18:52:39 (4 years ago)
Author:
kradhakrishnan
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • docs/CommissioningPlan/SDC4

    v1 v1  
     1== SDC4 Thermal background ==
     2•       **Phase**               :       Any 
     3•       **Status**              :       Draft   
     4•       **Category**            :       Daytime calibration (C3)
     5•       **Recurrence**  :       once every month starting from commissioning.
     6•       **Duration**            :       1 hour(s)
     7**Description**
     8•       Measure the instrument background and seasonal variations.
     9**Prerequisites**
     10**Procedure**
     11•       This calibration is performed by reading the detector without any light
     12•       CI BB_H, DIT=1800s, NDIT=2.
     13**Success criteria**
     14•       Accuracy of 1.2 e-/pix (FWHM of the distribution of a difference of backgrounds).
     15**Notes**
     16•       Same as activity 9.1 in AD4
     17**LBTO support**
     18
     19**Associated SHARK-NIR personnel**
     20
     21**Date performed and by whom**