    1515- Use the calibration flat-field source.
    1616- This calibration is performed illuminating the detector uniformly using the instrument FF lamp, for each filter.
    17 - It is expected that the detector FF depends on filters, final accumulated charge in the pixels. Thus, a series of exposures will be taken at different count levels (realized via ND filters and/or DIT) for each filter setup. In this case, the template takes the background frames automatically. Comparing different count rates at constant DIT will also yield a linearity map of the detector.
     17- It is expected that the detector FF depends on filters, the final accumulated charge in the pixels. Thus, a series of exposures will be taken at different count levels (realized via ND filters and/or DIT) for each filter setup. In this case, the template takes the background frames automatically. Comparing different count rates at constant DIT will also yield a linearity map of the detector.
    1818
    1919**Templates**
     20- SHARKNIR_gen_cal_InstrumentFlatField (not released)
    2021