== P3.4 Pupil alignment == • **Phase** : 3 • **Status** : Draft • **Category** : Nighttime testing (functionality and operability) (C5) + Nighttime calibration (C6) • **Recurrence** : before every observing target (CI, LSS), starting from commissioning. • **Duration** : 2 hours **Description** • Alignment of the internal pupil image with apodizer (at PP1) or Lyot stop (at PP2) in the coronagraphic mode **Prerequisites** • Bad pixel map, Background or Dark, Flat-field, detector RON **Procedure** • The position of the first tip-tilt mirror (IN_TT) is optimized in order to have the first pupil image centered with the apodizer/Lyot stop. This is obtained comparing the pupil image, formed on the SCICAM, with the image of the apodizer, obtained on the SCICAM using the calibration fiber as a source. **Success criteria** • Expected accuracy: 0.25% of pupil image diameter ~17arcsec of the TTM ~1.5pixel SCICAM ~30μm at the level of the SCICAM. **Notes** • Same as activity 8.2 in AD4 • Template: SHARKNIR_gen_tec_PupilAlignment **LBTO support** * AO support **Associated SHARK-NIR personnel** **Date performed and by whom** **Comment**