Changes between Version 8 and Version 9 of docs/CommissioningPlan/P3.4
- Timestamp:
- 02/02/21 13:06:24 (4 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
docs/CommissioningPlan/P3.4
v8 v9 6 6 • **Duration** : 2.5 hours 7 7 **Description** 8 • Alignment of the internal pupil image (PP1)with apodizer [and Lyot stop at PP2]in the coronagraphic mode8 • Alignment of the internal pupil image with apodizer (at PP1) or Lyot stop (at PP2) in the coronagraphic mode 9 9 **Prerequisites** 10 10 • Bad pixel map, Background or Dark, Flat-field, detector RON 11 11 **Procedure** 12 • The position of the first tip-tilt mirror (IN_TT) is optimized in order to have the first pupil image centered with the apodizer . This is obtained comparing the pupil image, formed on the SCICAM, with the image of the apodizer, obtained on the SCICAM using the calibration fiber as a source.12 • The position of the first tip-tilt mirror (IN_TT) is optimized in order to have the first pupil image centered with the apodizer/Lyot stop. This is obtained comparing the pupil image, formed on the SCICAM, with the image of the apodizer, obtained on the SCICAM using the calibration fiber as a source. 13 13 **Success criteria** 14 14 • Expected accuracy: 0.25% of pupil image diameter ~17arcsec of the TTM ~1.5pixel SCICAM ~30μm at the level of the SCICAM.