Changes between Version 18 and Version 19 of docs/CommissioningPlan/P3.4
- Timestamp:
- 03/03/23 14:27:50 (21 months ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
docs/CommissioningPlan/P3.4
v18 v19 14 14 15 15 **Procedure** 16 The internal deformable mirror's position (TTM_DM) is optimized to have the PSF image on the intermediate focal plane centered with the occulter. We compare the PSF image, formed on the SCICAM, with the image of the occulter obtained on the SCICAM using the calibration flat-field source.16 The internal deformable mirror's position (TTM_DM) is optimized to have the PSF image on the intermediate focal plane centered with the occulter. We compare the PSF center, measured on the SCICAM, with the center of the occulter measured on the SCICAM using the calibration flat-field source. 17 17 18 18 **Templates** … … 21 21 22 22 **Success criteria** 23 - the expected accuracy of 0.2 pixels (DV: need to check if we can reach this).23 - delta between PSF center and occulter center < 0.2 pixels. 24 24 25 25 **Notes**