== P3.22 Instrument flat field, detector linearity and bad pixels == • **Phase** : 3 • **Status** : Draft • **Category** : Daytime calibration (C3) • **Recurrence** : every day, with setup on request. • **Duration** : 20 minutes per configuration **Description** • Instrument flat field, detector linearity, and bad pixels. **Prerequisites** **Procedure** • Use the calibration flat-field source. • This calibration is performed illuminating the detector uniformly using the instrument FF lamp, for each filter. • It is expected that the detector FF depends on filters, final accumulated charge in the pixels. Thus, a series of exposures will be taken at different count levels (realized via ND filters and/or DIT) for each filter setup. In this case, the template takes the background frames automatically. Comparing different count rates at constant DIT will also yield a linearity map of the detector. **Success criteria** • 95% luminous uniformity at large scale. • 99% luminous uniformity at pixel-to-pixel scale 10-2. • Large scales filtered out, only small detector scale will remain. Effect of the large-scale flat field on DBI (rescaling in flux & wavelength) has negligible impact on DBI performances. **Notes** • Same as activity 7.9 in AD4 **LBTO support** • Support from LBTI/LBTO? **Associated SHARK-NIR personnel** • JF, LM, MdP **Date performed and by whom**