wiki:docs/CommissioningPlan/P3.15

Version 2 (modified by kradhakrishnan, 4 years ago) (diff)

--

P3.15 Flux calibration (off-centered target observation)

  • Phase : 3
  • Status : Draft
  • Category : Nighttime Science Calibration (C5)
  • Recurrence : once at commissioning and then on request for deep coronagraphic observations.
  • Duration : 20 minutes

Description

  • Flux calibration to allow contrast estimation between the central star and a close companion, by measuring the star flux out of the coronagraph

Prerequisites

  • Authorize SHARK-NIR on the SX side.
  • Atmospheric conditions: AO working conditions.
  • Bad pixel map, Background or Dark, Flat-field, distortion map

Procedure

  • On target, the same set-up as an observation on science target; AO closed-loop, ND for bright targets to prevent detector saturation, the star behind coronagraph centered and off-centered keeping the stop. The derotator mode should be tracking the field with possible adjustment of the angle to prevent the star from rotating around the coronagraph.
  • Observe a star with the coronagraphic devices, after an offset of 0.5” is applied.

Success criteria

  • Flux: 0.1 mag

Notes

  • Same as activity 7.2 in AD4

LBTO support

  • Support from LBTI/LBTO?

Associated SHARK-NIR personnel

  • JF, LM, MdP

Date performed and by whom