== P2.8 Looking for filter defects/artifacts + Science detector calibrations repeat == • **Phase** : 2 • **Status** : Draft • **Category** : Daytime testing (functionality and operability) (C2) • **Recurrence** : once at commissioning • **Duration** : 2 days **Description** • Finding the filter defects/artifacts using an internal calibration source (flat /on-axis PSF?) within the instrument. **Prerequisites** • NCPA measurement (within SHARK-NIR bench using internal sources) available. **Procedure** • Take the images of all the filters using the internal calibration source within the instrument to locate/find out the filter effects/artifacts. • You may take short- and long-exposures to get better contrast of the features if required. • Repeat the science calibrations below (some of these are performed when SHARK-NIR was in LBT cleanroom. Check [wiki:/docs/CommissioningPlan/P1.9 P1.9]) for details. ||SDC1||[wiki:/docs/CommissioningPlan/SDC1 Instrumental background for science]|| ||SDC2||[wiki:/docs/CommissioningPlan/SDC2 Instrument flat field, detector linearity and bad pixels]|| ||SDC3||[wiki:/docs/CommissioningPlan/SDC3 Wavelength calibration]|| ||SDC4||[wiki:/docs/CommissioningPlan/SDC4 Thermal background]|| ||SDC5||[wiki:/docs/CommissioningPlan/SDC5 Detector gain]|| ||SDC6||[wiki:/docs/CommissioningPlan/SDC6 Detector persistence]|| ||SDC7||[wiki:/docs/CommissioningPlan/SDC7 Spectral resolution]|| ||SDC8||[wiki:/docs/CommissioningPlan/SDC8 Ghost calibration]|| * Perform the working of this monitoring activity [wiki:/docs/Commissioning/SNM1 SNM1]. **Success criteria** • All the filter images saved. **Notes** • The time dedicated to this activity has to be considered in combination with the next activity if needed. Parallelism is possible. **LBTO support** **Associated SHARK-NIR personnel** **Date performed and by whom** **Comments** MB: I wonder why this needs to happen in this phase. I would say it could be done at any moment with internal sources, except of course for the dichroic / LBT mirrors [since we might need to apply the NCPA. after talking to EC and DV, I removed the use of RR / argos arm sicen it can all be done within shark-nir]