== [[span(style=color: green, **P2.3 Basic functionality, Instrument Operability and Functioning** )]] == - **Phase** : 2 - **Status** : [[span(style=color: green, **COMPLETE** )]] - **Category** : Daytime testing (functionality and operability) (C3) - **Recurrence** : once after instrument installation at telescope platform - **Duration** : 1 day **Description** - Activating the instrument after the installation of SHARK-NIR at the telescope platform. Verification of the shutdown procedure. - The functionality and operability of individual components/sub-systems are tested. Verification of the internal alignment of the instrument is performed using the light source within. The final step is the performance of the system. **Prerequisites** - SHARK-NIR is installed on the LBT telescope platform. **Procedure for Basic Functionality** - Switch on all the cabinets, follow the instrument telemetry and see if the telemetry is logging. - Turn all the hardware subsystems turned on and start all the corresponding services. Check all the logs from the services. - Check the shutdown procedure. - Testing the electronics and SW rotation of the knob controlling the cool down process (LBTI side). **Procedure for Basic Functionality** - Initialize all the motors/sub-systems. - Check the functionality of all the motors (home position, named/nominal positions, and reaching software limits). - Check the working of all the light sources. - Using the on-axis diffraction-limited light source, check the internal alignment of the SHARK-NIR. - Check the relative positions of the coronagraphic masks. - Check for on-axis optical quality. **Templates** - SHARKNIR_gen_init - SHARKNIR_gen_park - SHARKNIR_gen_LampsOff - SHARKNIR_cal_tec_NCPA_PD_lab (Phase Diversity approach) - SHARKNIR_cal_tec_NCPA_trial (Trial & Error approach) - SHARKNIR_di_mon_MaskRepeatability **Success criteria** - Activating instrument successfully. - All systems are turned on, and the services are all running without any errors. - The shutdown procedure is checked and working as it should be. - Motors and light sources are working as they should. - The internal alignment is preserved. - Performance of the system is acceptable. **Notes** **LBTO support** - Testing the working of the knob (motor or manual) controlling the cooldown. **Date performed and by whom**