== P1.9 Science detector calibrations == • **Phase** : 1 • **Status** : Draft • **Category** : Installation, integration, verification (C1) • **Recurrence** : once • **Duration** : 6 hour(s) **Description** • Calibration of the SHARK-NIR science detector: Gain, bad pixels, read noise, linearity, dark current, flat field, cross-talk, persistence, pattern noise. **Prerequisites** • SHARK-NIR connected to LBT network. • Electronics cabinets on. • The cryostat should be running and cooled down. • Internal alignment of the SHARK-NIR is verified. **Procedure** • check linearity, create bad pixel mask, bias calibration, master dark, and then perform the following daytime calibration procedures below. ||P3.20||[wiki:/docs/CommissioningPlan/P3.20 Instrumental background for science]||Day||(TBD)||C3||-|| ||P3.22||[wiki:/docs/CommissioningPlan/P3.22 Instrument flat field, detector linearity and bad pixels]||Day||2 hours||C3||-|| ||P3.23||[wiki:/docs/CommissioningPlan/P3.23 Wavelength calibration]||Day||2 hours||C3||-|| ||P3.24||[wiki:/docs/CommissioningPlan/P3.24 Thermal background]||Day||2 hours||C3||-|| ||P3.25||[wiki:/docs/CommissioningPlan/P3.25 Detector gain]||Day||2 hours||C3||-|| ||P3.26||[wiki:/docs/CommissioningPlan/P3.26 Detector persistence]||Day||2 hours||C3||-|| ||P3.27||[wiki:/docs/CommissioningPlan/P3.27 Spectral resolution]||Day||2 hours||C3||-|| ||P3.28||[wiki:/docs/CommissioningPlan/P3.28 Ghost calibration]||Day||2 hours||C3||-|| ||P3.30||[wiki:/docs/CommissioningPlan/P3.30 Filter wheels reproducibility monitoring]||Day||2 hours||C3||-|| **Success criteria** • linearity as expected • bad pixels mask created • bias and flat took **Notes** • Save all the intermediate images for Dino to check the SHARK-NIR Data Reduction pipeline. **LBTO support** **Associated SHARK-NIR personnel** • JF, LM, MdP, MM (from AZ), Dino? **Date performed and by whom**