== P1.9 Testing of SCICAM image grabbing by SOUL-AO == - **Phase** : 1 - **Status** : Draft - **Category** : Daytime testing (functionality and operability) (C3) - **Recurrence** : once - **Duration** : 0.5 days **Description** - Testing the SCICAM imaging grabbing (from a repository/folder) by SOUL-AO which is a pre-requisite for the SOUL-AO optimization for SHARK-NIR. **Prerequisites** - ICE compatibility between SHARK-NIR and SOUL-AO system. - SCICAM image should have a header following FITS standard. **Procedure** - Save a SCICAM image (with a PSF image) on the agreed folder or repository. - Check if SOUL-AO can grab the image and analyze the PSF. **Templates** **Success criteria** - SOUL-AO SW can grab the SCICAM image. **Notes** - The time dedicated to this activity has to be considered in combination with the other activities if needed. Parallelism is possible. - Points of contact: Alfio and Fabio from Arcetri. **LBTO support** - Liquid Nitrogen refill is required in the cleanroom. - Support of an IT personnel (for help related to network configuration) and an SW personnel (for help related to the TCS interface). **Date performed and by whom**