== P1.11 Check the working of the internal NCPA routine and looking for filter defects/artifacts == • **Phase** : 2 • **Status** : Draft • **Category** : Daytime testing (functionality and operability) (C2)??? • **Recurrence** : once at commissioning • **Duration** : 1 day **Description** • Check the working of the internal NCPA routine. Finding the filter defects/artifacts using an internal calibration source (flat /on-axis PSF?) within the instrument. **Prerequisites** • SCICAM cooled down. **Procedure** • Run the Phase-diversity routine using the images taken with the SHARK-NIR internal calibration sources to measure the NCPA within (excluding the deployable arm) * Take the images of all the filters using the internal calibration source within the instrument to locate/find out the filter defects/artifacts, if any. • You may take short- and long-exposures to get better contrast of the features if required. **Success criteria** * Internal NCPA routine works fine and the PSF shows improvement in its quality. • All the filter images saved. **Notes** • The time dedicated to this activity has to be considered in combination with the other activities if needed. Parallelism is possible. **LBTO support** **Associated SHARK-NIR personnel** **Date performed and by whom** **Comments** DM: flat field or on-axis fiber? I don't know how they do it in SPHERE, but I think a flat field source is more suitable.