Changes between Version 7 and Version 8 of docs/CommissioningPlan/P1.10
- Timestamp:
- 02/04/21 16:52:43 (4 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
docs/CommissioningPlan/P1.10
v7 v8 17 17 * Check the CRED-2 and ALPAO-DM SW and optical feedback. 18 18 * Take the images of all the filters using the internal calibration source (flat-field) within the instrument to locate/find out the filter defects/artifacts, if any. You may take short- and long-exposures to get better contrast of the features if required. 19 * Check the filter wheel position monitoring script detailed in [wiki:/docs/Commissioning/SNM1 SNM1]. Save the data in the repository. 19 20 20 21 **Success criteria** 21 22 • All the motors can reach the named positions / nominal positions. 22 23 • All the light sources are working and confirmed through optical feedback at the SCICAM. 24 * Filter wheel position monitoring script is run successfully and data saved in the repository. 23 25 * CRED-2 AND ALPAO-DM working normally. 24 26 • All the filter images saved.