Changes between Version 5 and Version 6 of docs/CommissioningPlan/P1.10


Ignore:
Timestamp:
02/03/21 17:49:41 (4 years ago)
Author:
kradhakrishnan
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • docs/CommissioningPlan/P1.10

    v5 v6  
    1 == P1.10 Testing of SCICAM image grabbing by SOUL-AO ==
     1== P1.10        Verification of the operability of the motors & light sources and looking for filter defects/artifacts ==
    22•       **Phase**               :       1 
    3 •       **Status**              :       Draft
     3•       **Status**              :       Draft   
    44•       **Category**            :       Unpack, Assemble, and Verification (C1)
    55•       **Recurrence**  :       once
    6 •       **Duration**            :       2 days
     6•       **Duration**            :       1.5
    77**Description**
    8 •       Testing the SCICAM imaging grabbing (from a repository/folder) by SOUL-AO which is a pre-requisite for the SOUL-AO optimization for SHARK-NIR
     8•       Checking the working of motors and light sources within SHARK-NIR using optical feedback. Finding the filter defects/artifacts using an internal calibration source (flat-field) within the instrument.
    99**Prerequisites**
    10 •       ICE compatibility between SHARK-NIR and SOUL-AO system
    11 * SCICAM image should have a header following FITS standard
     10•       SHARK-NIR connected to LBT network.
     11•       Electronics racks on.
     12•       DM best shape uploaded (shape minimizing internal aberrations)
     13•       The cryostat should be running and cooled down.
    1214**Procedure**
    13 •       Save a SCICAM image (with a PSF image) on the agreed folder or repository
    14 •       Check if SOUL-AO can grab the image and analyze the PSF
     15•       When the services are up and running, after the motors are homed, each of the individual motors is individually tested for their different named/nominal positions.
     16•       The light sources within SHARK-NIR are used simultaneously to check the working of the motors looking at the optical feedback.
     17* Check the CRED-2 and ALPAO-DM SW and optical feedback.
     18* Take the images of all the filters using the internal calibration source (flat-field) within the instrument to locate/find out the filter defects/artifacts, if any. You may take short- and long-exposures to get better contrast of the features if required.
     19
    1520**Success criteria**
    16 •       SOUL-AO SW can grab the SCICAM image
     21•       All the motors can reach the named positions / nominal positions.
     22•       All the light sources are working and confirmed through optical feedback at the SCICAM.
     23* CRED-2 AND ALPAO-DM working normally.
     24•       All the filter images saved.
     25
    1726**Notes**
    1827
    19 * Points of contact: Alfio and Fabio from Arcetri
    20 
    2128**LBTO support**
    22 
    23 * LBTO SW support may be required
    2429
    2530**Associated SHARK-NIR personnel**
    2631
    2732**Date performed and by whom**
     33
     34
     35**Comments**
     36* KR: add the filter monitoring routine check.